| Catalog Number |
253154000 |
| Medical Device Problem Codes |
Break (1069); Material Deformation (2976); Naturally Worn (2988); Scratched Material (3020)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/28/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that the head trials were worn, scratched and chipped.It did not happen in surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary - it was reported that top layer of tray was cracked.T handle was cracked.All head trials were worn, scratched and chipped.It did not happen in surgery.The product was returned to j&j medtech orthopaedics for evaluation.Visual inspection revealed deep scratches around the device's surface.Additionally, device exhibited signs of heavy usage.However, no signs of a stripped condition nor breakage could be identified over the device surface.Scratches observed were consistent with other tools and hard edges coming in contact with the device.Properly handling and attention to the approved use of the device diminishes the risk of failure.A dimensional inspection was not performed since it was not applicable to the complaint condition.A functional test was not performed since it was not applicable to the complaint condition.A device history record (dhr) review or manufacturing records evaluation (mre) was not possible because the date code j0302 provided is not a valid finished goods lot number.The overall complaint was confirmed as the observed condition of the articul tr ball grvd 36 +12 would have contributed to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot - a device history record (dhr) review or manufacturing records evaluation (mre) was not possible because the date code j0302 provided is not a valid finished goods lot number.H11 additional narrative: added: d4 lot, d9, h4, h6 medical device problem code corrected: d4 primary udi number,.
|
| |
|
Search Alerts/Recalls
|