|
Additional information: d9, g3, h3, h6.Complaint allegation is confirmed.One unpackaged and 4 sealed, ar-2324psp, serial/batch number (b)(6), were received for investigation.Visual inspection of the opened, damaged device, noted that the anchor was damaged at the distal end, the eyelet is still attached to the inserter.The o-ring and sutures were also returned.Functional testing was not performed due to the damage of the device.The most likely cause for the reported failure is attributed to misuse due to prying/leveraging the device during use.
|