|
Additional information: d9, g3, h3, h6.Complaint allegation is confirmed.Visual inspection noted a burn mark on the tip of the device.Functional testing could not be performed due to the damage to the tip.Per dfu-0242-7 at revision 0.E.Precautions.7.Always keep the tip of the active rf probe completely immersed in conductive irrigation fluid (saline, ringers¿ lactate, etc.), regardless of the type of arthroscopic surgery being performed.Do not use pre-warmed conductive irrigation fluids as this may result in tissue damage or thermal injury.8.A continuous flow of irrigation fluid is necessary during use of the apollo probes.Fluid flow assists in removing tissue debris and reducing the intra-joint temperature between activations.Rf probes utilized in stagnant fluid can result in heated irrigant in the joint, which can result in tissue damage inside the joint, skin burns near portals or result in damage to the probe.13.Prolonged ablation should be avoided.Intermittent pauses in ablation are recommended, to allow warm fluid and tissue debris to be evacuated from the joint.Prolonged probe activation while using the suction feature may result in elevated shaft or suction tubing temperatures.The most likely cause for this failure can be attributed to misuse due to the use of the device for extended periods, and/or excessive activation time of the device can cause it to overheat.
|