This report is being supplemented to provide additional information based on the approved final investigation.The device was returned to olympus for inspection, and the reportable malfunction was confirmed.Based on the results of the investigation, and although we could not specify the root cause of the suggested event, it is likely the defect was caused by physical stress to the distal end such as hitting or dropping and/or chemical stress such as medical solutions used for the device.A review of the device history record found no deviations that could have caused or contributed to the reported issue.The event can be prevented by following the instructions for use: instruction manual ltf-190-10-3d operation manual important information ¿ please read before use: contraindications, warnings, and precautions: do not strike, hit, or drop the endoscope¿s distal end, rigid portion, bending section, control section, universal cord, video connectors, or light guide connector.Also, do not bend, pull, or twist the endoscope¿s distal end, rigid portion, bending section, control section, universal cord, video connectors, or light guide connector with excessive force.The endoscope may be damaged and could cause patient injury, burns, bleeding, and/or perforations.It could also cause parts of the endoscope to fall off inside the patient.Instruction manual ltf-190-10-3d reprocessing manual 3.1 compatibility summary: methods listed as ¿compatible¿ in table 3.1 and 3.2 are compatible for routine use only when used according to manufacturer¿s instructions.Repeated use and reprocessing of endoscopes and accessories lead to gradual wear and tear.Furthermore, reprocessing methods that employ higher temperatures and more caustic/corrosive materials may lead to faster deterioration.In general, sterilization processes are harsher on equipment than disinfection processes.Before each patient procedure, inspect the endoscope and accessories for damage, according to the instructions described in this manual and its companion ¿operation manual.¿ reprocessing manual: 3.1 compatibility summary: list of compatible methods validated in terms of material durability: sterrad® 50/100s/200/nx¿: sterilization by sterrad® 50/100s/200/nx¿ system may deteriorate the adhesive of the insertion section.Depending on the circumstances, replacement of the insertion section may be required.Before use, confirm that the endoscope is free from any damage or other irregularities.For further details, contact olympus.Olympus will continue to monitor field performance for this device.Additional information added to fields d9, h2, h3, h4 and h6.
|