• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 3015516266 QUICKSET ACE GRATER CS COMPL; HIP INSTRUMENTS : INSTRUMENT CASES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 3015516266 QUICKSET ACE GRATER CS COMPL; HIP INSTRUMENTS : INSTRUMENT CASES Back to Search Results
Catalog Number 244000501
Medical Device Problem Codes Peeled/Delaminated (1454); Material Deformation (2976); Naturally Worn (2988)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/21/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the trays were found old and deteriorating.It did not happen in surgery.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H11 additional narrative: added: d9, d10 if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint #(b)(4) investigation summary
=
> it was reported that the trays found old and deteriorating.It did not happen in surgery.The product was returned to j&j medtech orthopaedics for evaluation.Visual analysis of the returned device revealed that the lid was missing.Signs of wear were observed on the overall surface, consistent with extensive and repeated use over many years.There was no evidence of peeling coating.The device also exhibited surface discoloration, which was identified as an indicator of end of life condition.The lifecycle requirements of the device are event related and depend on the use and inspection of the device in clinical practice.As the device can be damaged on the first or 100th use, the device must be properly inspected prior to each surgical use.Refer to the device/country specific ifu for information related to end of life, reprocessing instructions, and inspection procedures.A functional test was not performed since it was not applicable to the complaint condition.A dimensional inspection was not performed since it was not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the quickset ace grater cs compl would have contributed to the complained issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot
=
> the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.H11 additional narrative: corrected: h3.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUICKSET ACE GRATER CS COMPL
Common Device Name
HIP INSTRUMENTS : INSTRUMENT CASES
Manufacturer (Section D)
DEPUY IRELAND - 3015516266
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key20579901
Report Number1818910-2024-22832
Device Sequence Number8650313
Product Code FSM
UDI-Device Identifier10603295123712
UDI-Public(01)10603295123712(10)CR6BW4000
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/31/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number244000501
Device Lot NumberCR6BW4000
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/29/2024
Supplement Date Received by Manufacturer11/04/2024
11/08/2024
Initial Report FDA Received Date10/31/2024
Supplement Report FDA Received Date11/05/2024
11/11/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/02/2008
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
QUICKSET ACE GRATER CS B TRAY; QUICKSET ACE GRATER CS B TRAY; QUICKSET ACE GRATER CS B TRAY; QUICKSET ACE GRATER CS B TRAY; QUICKSET ACE GRATER CS COMPL; QUICKSET ACE GRATER CS COMPL; QUICKSET ACE GRATER CS COMPL; QUICKSET ACE GRATER CS COMPL; QUICKSET ACE GRATER CS M TRAY; QUICKSET ACE GRATER CS M TRAY; QUICKSET ACE GRATER CS M TRAY; QUICKSET ACE GRATER CS M TRAY; QUICKSET ACE GRATER CS T TRAY; QUICKSET ACE GRATER CS T TRAY; QUICKSET ACE GRATER CS T TRAY; QUICKSET ACE GRATER CS T TRAY
Patient SexUnknown
-
-