|
RESPIRONICS, INC. BIPAP A40; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
|
Back to Search Results |
|
| Model Number |
R1076495 |
| Medical Device Problem Code |
Electrical /Electronic Property Problem (1198)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/07/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The manufacturer received information alleging a ventilator inoperative condition occurred.There was no harm or injury reported.The device was sent to a service center for evaluation.During evaluation and review of the device error logs, it was identified that an error was logged which indicates that a log was not written correctly.The ventilator inoperative condition was not duplicated.The device's main board was replaced to prevent reoccurrence.The unit was confirmed to be working properly after replacement.Prior to identifying the issue described in fsn 2023-cc-src-039, complaints associated with the issue did not meet requirements for vigilance reporting as a reportable incident.As part of the hhe process, a retrospective review of complaints associated with the issue described in fsn 2023-cc-src-039 was necessary to reassess reportability.Upon this further assessment, these complaints are being reported.
|
| |
|
Search Alerts/Recalls
|
|
|