| Catalog Number |
45-650 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
07/25/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Healthcare professional reported, left side rupture.Device explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side rupture.Device explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: based on the product analysis performed, the assessments of the complaints are: rupture: observed a broken assessed as fold flaw opening and observed missing shell assessed as inconclusive.As per the investigation procedure, creases and non-penetrating nick were completed and none of the observations are found to be potentially related to the manufacturing process, no further actions are required.Additional, changed, and/or corrected data: d9; h3; h6.
|
| |
|
Search Alerts/Recalls
|