Catalog Number 33310C
Medical Device Problem Codes Fracture (1260); Detachment of Device or Device Component (2907)
Health Effect - Clinical Codes Foreign Body In Patient (2687); Insufficient Information (4580)
Date of Event 10/14/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that tip of the bowel grasper snapped off during use inside a patient.Caused the procedure to be prolonged by 3 hours while retrieving small bit that had come off.Since the procedure was prolonged by 3 hours and no information if the broken part was fully retrieved, this case is reportable as precaution.Cross reference mdr: 9610617-2024-00429.
 
Additional Manufacturer Narrative
The affected device has been requested for investigation by the manufacturer.The device was not yet returned for investigation.Cross reference case id: (b)(4).The event is filed under internal karl storz complaint id: (b)(4).
 
Additional Manufacturer Narrative
The investigation shows that, based on the customer's description, user error is the likely cause.Compared to previous complaints, it appears that a branch of the mouthpiece broke, potentially due to excessive force.Additionally, the unknown age of the insert may have contributed to material breakage.However, as no article was provided, a detailed investigation could not be conducted.The diagnosed issue is not caused by a production defect.The event is filed under internal karl storz complaint id: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLICKLINE FORCEPS INSERT
Common Device Name
RIGID ENDOSCOPIC TISSUE MANIPULATION
Manufacturer (Section D)
KARL STORZ SE & CO. KG
dr,-kark - storz-strasse 34
tuttlingen, 78532
GM  78532
Manufacturer (Section G)
KARL STORZ SE & CO. KG
dr,-kark - storz-strasse 34
tuttlingen, 78532
GM   78532
Manufacturer Contact
anja fair
2151 e grand ave
el segundo, CA 90245
MDR Report Key20583043
Report Number9610617-2024-00432
Device Sequence Number8171169
Product Code GCJ
UDI-Device Identifier04048551134092
UDI-Public04048551134092
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K935071
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/03/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number33310C
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/15/2024
Supplement Date Received by Manufacturer11/15/2024
Initial Report FDA Received Date10/31/2024
Supplement Report FDA Received Date12/03/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-