| Model Number |
INGENIA 1.5T CX |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/28/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The strip at the opening of the magnet that has come loose was secured again and the system was handed back to the customer for use.The failure of the qbc seal to remain in place is considered a malfunction.Remedial action: a field action was initiated under 2023-pd-mr-013.A field safety notification was send to the customers informing them about this potential issue.A field correction is scheduled to be released in q4-2024.
|
| |
|
Event or Problem Description
|
|
Philips received a report that the strip at the opening of the magnet has come loose.This seal is used for protection, to avoid the patient from potentially contacting the sharp edges of the magnet covers.If this seal is damaged, it is likely that this could lead to serious injury.
|
| |
|