• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. COMP LK SCR 3.5HEX 4.75X15 ST; SHOULDER PROSTHESIS, REVERSE CONFIGURATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. COMP LK SCR 3.5HEX 4.75X15 ST; SHOULDER PROSTHESIS, REVERSE CONFIGURATION Back to Search Results
Model Number N/A
Medical Device Problem Codes Detachment of Device or Device Component (2907); Unintended Movement (3026)
Health Effect - Clinical Code Inadequate Osseointegration (2646)
Date of Event 10/04/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
This event is recorded by zimmer biomet under (b)(4).G2, country event occurred in: japan.The product is in the process of being evaluated by zimmer biomet.Once the product investigation has been completed, a follow up/final report will be submitted.
 
Event or Problem Description
It was reported that that the patient was revised due to disassociation.It was also reported that an unknown amount of screws were loose from the base plate.Due diligence is complete.No additional information is available.
 
Additional Manufacturer Narrative
This event is recorded by zimmer biomet under (b)(4).The following sections have been updated: b4, b5, g1, g3, g6, h2, h3, h6, h11.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history record identified no deviations or anomalies during manufacturing.Radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: image demonstrates interval loosening and displacement of the glenoid implant.The glenosphere appears disassociated from the baseplate with resulting glenohumeral subluxation.A definitive root cause cannot be determined.The reported event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No additional information is available regarding the incident.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMP LK SCR 3.5HEX 4.75X15 ST
Common Device Name
SHOULDER PROSTHESIS, REVERSE CONFIGURATION
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key20587294
Report Number0001825034-2024-02574
Device Sequence Number6529052
Product Code PHX
UDI-Device Identifier00880304677135
UDI-Public(01)00880304677135(17)340427(10)66714691
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K130390
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/30/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number180550
Device Lot Number66714691
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/07/2024
Supplement Date Received by Manufacturer04/24/2025
Initial Report FDA Received Date11/01/2024
Supplement Report FDA Received Date04/30/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/27/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexFemale
-
-