This event is recorded by zimmer biomet under (b)(4).The following sections have been updated: b4, b5, d4, g1, g3, g6, h2, h3, h4, h6, h11.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history record identified no deviations or anomalies during manufacturing.Radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: image demonstrates interval loosening and displacement of the glenoid implant.The glenosphere appears disassociated from the baseplate with resulting glenohumeral subluxation.A definitive root cause cannot be determined.The reported event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|