• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. ADVANCED PATELLA 32MM 3 PEG IMPLANT; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. ADVANCED PATELLA 32MM 3 PEG IMPLANT; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Catalog Number 200-07-32
Medical Device Problem Codes Naturally Worn (2988); Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 09/23/2020
Type of Reportable Event Serious Injury
Event or Problem Description
Legal case ¿ usa.Original description summary: as reported via legal documentation the patient had a left knee replacement.Approximately 2 years and 10 months after the initial procedure the patient had a left knee revision on (b)(6) 2020.There is no other patient demographic or medical history available.There is no information on the surgical procedure or patient outcome.There is no device return.There are no photos or other images of the device provided.No additional information is available.Unable to determine right and left products.The patient has filed a complaint in a coordinated action in alachua county with master case no.(b)(6).The consolidated long form complaint that applies to cases filed in this coordinated action alleges that patients filing suits in this coordinated action were required ¿to undergo revision surgeries due to severe, pain, swelling, and instability¿ due to ¿wear of the polyethylene components and resulting component loosening and/or other failure failures causing serious complications including tissue damage, osteolysis, permanent bone loss, and other injuries.¿ because the patient has filed a short-form complaint in this coordinated action, the patient appears to allege that the patient was injured as a result of wear of an exactech polyethylene device.Additional information received from legal on 12sep2023 as reported via legal documentation the patient had a left knee replacement.Approximately 2 years and 10 months after the initial procedure the patient had a left knee revision on (b)(6) 2020.Revision op report postoperative diagnosis: aseptic loosening, left femur and patella.There was some erosion under the medial aspect of the patella and it was loose.The femoral component was checked by tapping on the anterior flange and it came loose from the cement.Cement mantle was intact.The well-fixed tibial component was removed with minimal additional bone loss.There is no other patient demographic or medical history available.There is no device return.There are no photos or other images of the device provided.No additional information is available.
 
Additional Manufacturer Narrative
1038671-2023-02331 d10: concomitant devices: truliant psc tibial insert (02-022-35-3510, 5060501).Truliant fit tibial tray (02-022-45-3535, 4856602).Truliant femoral component (02-020-11-0235, 5046120).Manufacturer narrative: the reason for the revision reported cannot be determined from the available information but may have been reported at this later time due to inclusion of one of the implanted devices in the packaging recall.However, the reported prosthesis wear/failure, femoral loosening and patellar loosening could not be confirmed and potential contributions of user, or patient-related considerations to the event could not be assessed as no images or radiographs were provided, relevant clinical information was not provided, and the devices were not available for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCED PATELLA 32MM 3 PEG IMPLANT
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH INC.
2320 nw 66th court
gainesville FL 32653
Manufacturer Contact
geoff gannon
2320 nw 66th ct
gainesville, FL 32653
3523771140
MDR Report Key20587391
Report Number1038671-2024-04207
Device Sequence Number20470602
Product Code JWH
UDI-Device Identifier10885862314284
UDI-Public10885862314284
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K160484
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2020
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/28/2022
Device Catalogue Number200-07-32
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/31/2023
Initial Report FDA Received Date11/01/2024
Date Device Manufactured08/29/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age56 YR
Patient SexFemale
-
-