| Catalog Number |
00840004410 |
| Medical Device Problem Codes |
Adverse Event Without Identified Device or Use Problem (2993); Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Insufficient Information (4580)
|
| Date of Event |
01/30/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Multiple mdr reports were filed for this event.Associated reports are available for the applicable concomitant products below in the d10 narrative.D10: concomitant medical products, part (lot): - 00840009400 (b)(6).- 00840002511 (b)(6) associated product information: - 110034355 (b)(6) - 47840103800 (b)(6) - 47225500800 (b)(6) - 00840108000 (b)(6) - 00840009400 (b)(6) - 00840009500 (b)(6) - 00840009000 (b)(6).The customer has indicated that the product will not be returned to zimmer biomet for investigation, as it remains implanted.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.If any further information is found that would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient is considering an elbow revision procedure due to unknown reasons to remove the humeral stem implanted approximately four (4) years ago and other elbow components implanted approximately one and a half (1.5) years ago.This planned revision will involve replacements with competitor products on an unknown date.Attempts have been made, and no further information has been provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The following sections were updated: b1; b2; b4; b5; b7; d2; d6b; d10; e1; g1; g3; g6; h1; h2; h3; h6; h10.D10: concomitant medical products, part (lot): 00840002511 (63028261).Associated product information: 00840009400 (65385845), 110034355 (ax32bf1004), 47840103800 (64401469), 47225500800 (65882521), 00840108000 (56708036), 00840009400 (65385845), 00840009500 (65337404), 00840009000 (64302321).If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
It was reported a patient had a right total elbow arthroplasty approximately five (5) years ago followed by a revision forty-one (41) months later due to pain and ulnar loosening.Subsequently, one (1) month ago, the patient underwent a second revision due loosening of the ulnar and humeral implants.A competitor, custom made, ulnar implant and a zb humeral implant were placed.No further event information is available at the time of this report.
|
| |
|
Event or Problem Description
|
|
No further event information is available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.Proposed annex g code: mechanical (g04) - stem.The reported event is not confirmed.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.A review of the device history record(s) identified no deviations or anomalies during manufacturing related to the reported event.Medical records were provided related to the linked complaint and reviewed by a healthcare professional.A timeline was created for this patient.A definitive root cause cannot be determined.If any further information is found that would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Search Alerts/Recalls
|