• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. URETERO-RENO FIBERSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. URETERO-RENO FIBERSCOPE Back to Search Results
Model Number URF-P7
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Kidney or Urinary Problem (4503); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 05/21/2024
Type of Reportable Event Serious Injury
Event or Problem Description
Olympus medical systems corp.(omsc) reviewed a literature titled "predictive factors for prolonged operative time in ureteroscopic lithotripsy for ureteral stones".Literature summary introduction: a prolonged operative time of lithotripsy with ureteroscopy for urolithiasis increases the risk of infectious complications; however, few reports have investigated the factors prolonging the operative time for ureteral stones.We investigated the factors associated with longer operative time in ureteroscopy for ureteral stones.Methods: this retrospective cohort study analyzed patients who underwent retrograde ureteroscopic lithotripsy for ureteral stones and achieved an endoscopic stone-free status between april 2019 and july 2022.Patients were classified into two groups based on an operative time of =90 minutes or <90 minutes.We compared the patient and stone characteristics and surgical outcomes, and investigated the factors associated with a prolonged operative time.Results: the cohort comprised 519 patients, with 58 patients in the group with an operative time of =90 minutes.Compared to the shorter operative time group, the longer operative time group had a significantly greater proportion of males, stone diameter, stone volume, and hounsfield units of stone; additionally, the longer operative time group had higher prevalences of endoscopic findings of edema, polyps, and mucosa-stone adherence.Multivariable analysis showed that stone size >10 mm (odds ratio 4.05), polyps (odds ratio 2.40), and mucosal adherence (odds ratio 3.51) were significantly associated with an operative time exceeding 90 minutes.There were no significant differences between the two groups in the incidences of postoperative fever and systemic inflammatory response syndrome.Conclusions: stone size, endoscopic findings of polyps, and mucosa-stone adherence were independent factors associated with a longer operative time.Type of serious adverse events/number of patients sirs (systematic inflammatory response syndrome) (48 patients) ureteral injury (20 patients).
 
Additional Manufacturer Narrative
H6 health effect - clinical code 4581 - appropriate term / code not available is used to reflect sirs (systematic inflammatory response syndrome) h6 health effect - clinical code 4503 - unspecified kidney or urinary problem is used to reflect ureteral injury the evaluation of the event is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.Additionally, to provide an update to field (d8) and to provide a correction to field (d4).Since the serial and/or lot number was not provided, the (udi) unique device identifier is unknown and have been left blank accordingly.The device history record was unable to be reviewed for this device since the serial and/or lot number was not provided.However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.Based on the results of the investigation, the reported event was likely an accident, or a complication associated with a surgical procedure while using the subject device.There was no complaint reported on the subject device and no evidence of an olympus device malfunction.Therefore, the root cause cannot be determined.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
URETERO-RENO FIBERSCOPE
Common Device Name
URETERO-RENO FIBERSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key20588031
Report Number9610595-2024-22448
Device Sequence Number8804699
Product Code FGB
UDI-Device Identifier04953170435096
UDI-Public04953170435096
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K181451
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Study,Literature,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/10/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberURF-P7
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/08/2024
Supplement Date Received by Manufacturer12/26/2024
Initial Report FDA Received Date11/01/2024
Supplement Report FDA Received Date01/10/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
1.5 FRENCH NITINOL BASKET BY COOK MEDICAL; 4.5 FRENCH SEMI-RIGID URETEROSCOPE BY RICHARD WOLF; 6 FRENCH SEMI-RIGID URETEROSCOPE BY RICHARD WOLF; UNKNOWN ACCESS SHEATH; UNKNOWN HOLMIUM-YTTRIUM ALUMINUM GARNET LASER; UNKNOWN URETERAL STENT WITH STRING
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-