It was reported that, during a cori assisted tka surgery, one (1) real intelligence 5 mm cylindrical bu popped out right before burring and would not seat properly.The procedure was resumed, after a non-significant delay, using a s+n backup device.No injury was reported as a result of this issue.
|
|
Internal complaint reference: case (b)(4).H3, h6: the real intelligence 5 mm cylindrical bu, part number rob10035, lot number 51242294, used for treatment was not returned for evaluation, therefore a device analysis was unable to be performed.A complaint history review for similar reported/confirmed complaints concluded this was an isolated event.A review of manufacturing records indicate the device met all specifications upon release into distribution.The failure mode and associated risk have been anticipated within the risk file and the documented risk level is still adequate.A historical escalation event review was not completed.The product was not returned and no evidence was made available to link the complaint to an escalation event.We have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Although the reported problem was not confirmed through a visual or functional evaluation, factors contributing to the reported symptom may have been associated with product mishandling during surgical site assembly/disassembly or decontamination.Based on the investigation, no containment or corrective action is recommended or required at this time.Should any additional information be received the complaint will be reopened.The failure mode will continue to be closely monitored through complaint investigation and trended through post market surveillance activities.
|