ST. JUDE MEDICAL SWARTZ¿ BRAIDED TRANSSEPTAL GUIDING INTRODUCER, LAMP¿ 45, 63 CM LENGTH, 8F; INTRODUCER, CATHETER
|
Back to Search Results |
|
| Model Number |
407360 |
| Medical Device Problem Code |
Gas/Air Leak (2946)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
During a procedure, the sealing lip of the sheath was leaking, and air bubble entered the system.The sheath was replaced, and the procedure was completed with the patient remaining stable.
|
| |
|
Additional Manufacturer Narrative
|
|
One 8.5f swartz braided introducer sheath was received for evaluation.Functional testing confirmed an air and fluid leak in the hemostasis valve.The cap was removed from the hemostasis hub and the hemostasis seals were microscopically inspected.A puncture was noted in the proximal and distal seals.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the punctured seals and subsequent leak remains unknown.The ifu states: do not remove dilator or catheter rapidly.Damage to the valve may occur, potentially compromising hemostasis.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d4, g3, h2, h4.
|
| |
|