• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL CIRCULATION, INC. AUTOPULSE®RESUSCITATION SYSTEM; CARDIAC RESUSCITATOR, LINE-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL CIRCULATION, INC. AUTOPULSE®RESUSCITATION SYSTEM; CARDIAC RESUSCITATOR, LINE-POWERED Back to Search Results
Model Number 100
Medical Device Problem Code Visual Prompts will not Clear (2281)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/05/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The autopulse platform (sn (b)(6) associated with this complaint will not be returned for evaluation.However, the customer cleared the user advisory (ua) 45 (not at "home" position after power-on/restart) error message using the instructions provided by a zoll representative.The autopulse platform is functional and was placed back to clinical use.Note that the user advisory error messages are clearable by the user.Per the autopulse resuscitation system model 100 user guide, if the driveshaft is not at its home position when the autopulse is powered on, a user advisory (45) will occur.This user advisory will persist until the driveshaft is returned to its home position.To clear a user advisory (45) pull up on the lifeband until the chest bands are fully extended (thereby moving the driveshaft back to its home position), and then restart.
 
Event or Problem Description
During shift check, after the call, the autopulse platform (sn (b)(6) displayed user advisory (ua) 45 (not at "home" position after power-on/restart) when the user replaced the lifeband.The user was not able to clear the user advisory.No patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOPULSE®RESUSCITATION SYSTEM
Common Device Name
CARDIAC RESUSCITATOR, LINE-POWERED
Manufacturer (Section D)
ZOLL CIRCULATION, INC.
2000 ringwood ave.
san jose CA 95131
Manufacturer (Section G)
ZOLL CIRCULATION, INC.
2000 ringwood ave.
san jose CA 95131
Manufacturer Contact
kimthoa sackrison
2000 ringwood ave.
san jose, CA 95131
MDR Report Key20591305
Report Number3010617000-2024-00848
Device Sequence Number6529189
Product Code DRM
UDI-Device Identifier00849111001014
UDI-Public00849111001014
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K112998
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 11/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number100
Device Catalogue Number8700-0740-01
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/07/2024
Initial Report FDA Received Date11/01/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-