STRYKER ORTHOPAEDICS-MAHWAH TRIDENT PSL WITH PUREFIX HA 50MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
|
Back to Search Results |
|
| Catalog Number |
542-11-50E |
| Medical Device Problem Codes |
Loss of Osseointegration (2408); Malposition of Device (2616)
|
| Health Effect - Clinical Code |
Ambulation or Postural Difficulties (2544)
|
| Date of Event |
10/07/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Reported event: an event regarding loosening and malposition involving a trident shell was reported.The event was not confirmed.Method & results product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.Clinician review: no medical records were received for review with a clinical consultant.Product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: the exact cause of the event could not be determined because insufficient information was provided.Further information such as return of the device, pathology reports, pre- and post-operative x-rays and the primary operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and / or additional information become available to indicate further evaluation is warranted, this record will be reopened.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient's right hip was revised due to shell migration and loosening after the patient suffered a fall.The shell, a screw, head, and liner were revised.There were no allegations against the revised head or liner.Rep confirmed that no further information will be released by the hospital or surgeon.
|
| |
|
Search Alerts/Recalls
|
|
|