• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIFESCAN EUROPE GMBH OT VERIO FLEX METER; GLUCOSE MONITORING SYS/KIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIFESCAN EUROPE GMBH OT VERIO FLEX METER; GLUCOSE MONITORING SYS/KIT Back to Search Results
Lot Number 5813216
Medical Device Problem Code Device Displays Incorrect Message (2591)
Health Effect - Clinical Code Hypoglycemia (1912)
Date of Event 10/23/2024
Type of Reportable Event Serious Injury
Event or Problem Description
On (b)(6) 2024, a reporter for the lay user/patient contacted lifescan (lfs) united states, alleging that the patient¿s onetouch verio flex meter was displaying an ¿error 5¿ message during testing.The complaint was classified based on the customer care agent (cca) documentation.The reporter stated that on (b)(6) 2024, between 6:00 pm and 6:30 pm, the patient was feeling ¿loopy¿.The patient drank orange juice and ate peanut butter then attempted to test her blood glucose at 6:30 pm but the alleged error message began prompting.The emergency medical services (ems) were reportedly notified for assistance.The reporter claimed that when ems arrived at 7:00 pm, they tested the patient¿s blood glucose on the ems meter and obtained a reading of ¿42 mg/dl¿ and treated her with intravenous fluids.At the time of troubleshooting, the cca noted that the subject meter was not being used for the first time.The cca walked through resolving issue and the alleged issue was resolved.Replacement products were sent to the patient.This complaint is being reported because the patient reportedly received medical intervention for an acute low blood glucose excursion after the alleged meter issue began.The subject meter could not be ruled out as a contributor to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OT VERIO FLEX METER
Common Device Name
GLUCOSE MONITORING SYS/KIT
Manufacturer (Section D)
LIFESCAN EUROPE GMBH
dammstrasse 19
zug 6300
SZ  6300
MDR Report Key20593767
Report Number2939301-2024-00079
Device Sequence Number15645870
Product Code NBW
UDI-Device Identifier00353885010986
UDI-Public00353885010986
Combination Product (Y/N)N
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Importer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 11/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Lot Number5813216
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date10/28/2024
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date11/01/2024
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-