| Model Number |
SK51 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/15/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
It was reported that after an initial bunion surgery, all hardware was removed on (b)(6) 2024 due to loss of correction.There were no deficiencies or malfunctions alleged/found with any tmc device and no other patient outcomes were reported as a result of this event.Additional information indicates the patient had poor bone quality.No devices were returned for evaluation.Device-specific information was not available; therefore, a review of device history records was not able to be performed.However, all non-conformances for possible kits utilized in surgery were reviewed and no non-conformances or issues during the manufacture or release of the products were identified to date that could have contributed to what was reported.Although a number of factors could have contributed to what the patient experienced, the most likely cause cannot be determined based on the available information and without the return of the device.However, it is likely the patient's poor bone quality contributed to the reported event.The company will supplement this mdr as necessary and appropriate.Additional tmc devices explanted in the same revision surgery were reported in 3011623994-2024-00199.
|
| |
|
Event or Problem Description
|
|
It was reported that after an initial bunion surgery, all hardware was removed on (b)(6) 2024 due to loss of correction.There were no deficiencies or malfunctions alleged/found with any tmc device and no other patient outcomes were reported as a result of this event.
|
| |
|
Search Alerts/Recalls
|