| Catalog Number |
G310SCLR |
| Medical Device Problem Code |
Restricted Flow rate (1248)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
While using a cavitron 300 series g310, they allege that they have low water flow and the handpiece heated up, no injury resulted.
|
| |
|
Additional Manufacturer Narrative
|
|
There has been a previous report received where lack of water flow has caused an overheating insert.Since an overheating insert could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
|
| |
|
Additional Manufacturer Narrative
|
|
11/25/24 repair tech: (b)(4).W4e water sol,iso valve clogged.Water system was clogged causing poor waterflow.Which was making the tip get hot.Debris buildup in the water solenoid.Poor water flow regulation.Debris buildup in the water filter.Debris buildup in the handpiece cable.Replaced damaged/worn components and recalibrated unit to factory specs.Qa: (b)(4).
|
| |
|