The product investigation was completed.Device evaluation details: the device was returned to biosense webster (bwi) for evaluation.A visual inspection and screening test of the returned device was performed following bwi procedures.Visual analysis revealed a separation between pebax and electrode section, and a reddish material inside it.The device was connected to the carto 3 system, and no errors were observed.The force feature of the device was evaluated and the force values and the vector were observed within specifications.No force issues were observed.A manufacturing record evaluation was performed for the finished device number lot 31146531m and no internal action related to the complaint was found during the review.The reddish material inside the pebax could be related to the force issue reported by the customer; therefore, the customer complaint was confirmed.The potential cause of the pebax separation cannot be conclusively determined.The instructions for use (ifu) states: when cleaning the tip electrode, be careful not to twist the tip electrode with respect to the catheter shaft; twisting may damage the tip electrode bond and loosen the tip electrode, or may damage the contact force sensor.In addition, in order to prevent damage to the catheter tip, use the insertion tube supplied with the catheter to advance or retract the catheter through the hemostasis valve of the sheath.After insertion, slide the insertion tube back toward the handle.As part of biosense webster's quality process, all devices are manufactured, inspected, and released to approved specifications.This product issue will be addressed through bwi's quality system.An internal action was created for further investigation of the force sensor sleeve insolation to prevent fluids to get inside into the device.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
|
|
On 9-nov-2024, the product investigation was updated to include additional h section codes.Updates: - h6 investigation findings has had "manufacturing process problem identified (c16)".- h6 investigation conclusions has had "caused traced to manufacturing (d03)" applied.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
|