BARD PERIPHERAL VASCULAR, INC. ENCOR ENSPIRE SYSTEM; REUSABLE BIOPSY Back to Search Results
Catalog Number E4115
Medical Device Problem Codes Application Program Problem (2880); Unintended Movement (3026)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/20/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a biopsy procedure, a pattern was selected, however the unit ignored desired pattern and sampled a full clock rotation, and then device stopped responding.It was further reported that the touchscreen, driver and pedal were all allegedly stopped responding.It has been turned off and back on, driver and pedal detached and reattached and new needle loaded and still it does not respond.Furthermore, we had set the machine to avoid the 7:00 and 9:00 positions.Although the screen showed those positions grayed out, the needle did not avoid these areas when sampling.It ignored the pattern and took a full clock sample instead.After the samples were taken, the needle began pulse as if it was stuck and the unit stopped responding.We used the foot pedal to initiate sampling.After samples were taken and the unit became stuck and stopped responding, we attempted to use both the foot pedal and the sample button on the hand piece to reset the needle and change to marker position but were unsuccessful.We were able to obtain the samples needed, but the machine also sampled two positions that we did not intend to sample.The needle and unit stopped responding after samples were obtained, and we were allegedly unable to switch to the marker placement function.There was no reported patient injury.
 
Additional Manufacturer Narrative
H10: as the serial number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.H11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Additional Manufacturer Narrative
H10: manufacturing review: a dhr and manufacturing review was not required as the device was not returned.Investigation summary: the physical device was not returned for evaluation.No photos or videos were provided for review.Therefore, the investigation is inconclusive for the alleged issue as no objective evidence was provided for review.A definitive root cause for the reported failed to respond/unintended sampling issue could not be determined based upon the provided information.Labeling review: the labeling/packaging review was not required as the reported event is not associated with a labeling, packaging, or use related issue.H10: g3, h6 (method).H11: d4 (unique identifier (udi) #).H11: section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that during a biopsy procedure, a pattern was selected, however the unit ignored desired pattern and sampled a full clock rotation, and then device stopped responding.It was further reported that the touchscreen, driver and pedal were all allegedly stopped responding.It has been turned off and back on, driver and pedal detached and reattached and new needle loaded and still it does not respond.Furthermore, we had set the machine to avoid the 7:00 and 9:00 positions.Although the screen showed those positions grayed out, the needle did not avoid these areas when sampling.It ignored the pattern and took a full clock sample instead.After the samples were taken, the needle began pulse as if it was stuck and the unit stopped responding.We used the foot pedal to initiate sampling.After samples were taken and the unit became stuck and stopped responding, we attempted to use both the foot pedal and the sample button on the hand piece to reset the needle and change to marker position but were unsuccessful.We were able to obtain the samples needed, but the machine also sampled two positions that we did not intend to sample.The needle and unit stopped responding after samples were obtained, and we were allegedly unable to switch to the marker placement function.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENCOR ENSPIRE SYSTEM
Common Device Name
REUSABLE BIOPSY
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
DYMAX CORP. -2523003
110 marshall drive
warrendale PA 15086
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key20596959
Report Number2020394-2024-01648
Device Sequence Number6558680
Product Code KNW
UDI-Device Identifier00801741086113
UDI-Public(01)00801741086113
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
K111100
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/08/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberE4115
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/08/2024
Supplement Date Received by Manufacturer01/08/2025
Initial Report FDA Received Date11/04/2024
Supplement Report FDA Received Date01/22/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient Age65 YR
Patient SexFemale
Patient Weight89 KG
-
-