| Catalog Number |
7209685 |
| Medical Device Problem Code |
Overheating of Device (1437)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/23/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H10: internal complaint reference number: (b)(4).H3, h6: this complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known inherent device and/or procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
|
| |
|
Event or Problem Description
|
|
It was reported that during a shoulder surgery, the temperature of the saphyre probe was too high after using it for more than 10 minutes, the surgeon was afraid to use it anymore.The procedure was completed with a surgical delay of less than 30 minutes using a smith and nephew back up device.No further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: h2: corrected data on: d4 (lot number, expiration date (should be left in blank)) & h4 (should be left in blank).
|
| |
|
Additional Manufacturer Narrative
|
|
H10: internal complaint reference: (b)(4).H3, h6: the reported device was received for evaluation.A visual inspection was performed on the exterior of product and no physical damage was observed.Functional evaluation revealed no issues.The wand produced plasma as expected and did not show signs of running at higher power.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found no similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The root cause was not determined since the reported malfunction could not be duplicated during the product evaluation process.Factors that can contribute to the failure include procedural variance, or deviations from established protocols, increasing risks and leading to unintended consequences (improper cleaning and sterilization methods).Please refer to the instructions for use for precautions and warnings related to the use of the device.Based on this investigation, the need for corrective action is not indicated.
|
| |
|