• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. VERSA-DIAL/COMP TI STD TAPER; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER, UNCEMENTED/EXTREMITIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. VERSA-DIAL/COMP TI STD TAPER; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER, UNCEMENTED/EXTREMITIES Back to Search Results
Catalog Number 118001
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Joint Dislocation (2374)
Date of Event 10/07/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient underwent an initial shoulder arthroplasty on and unknown date.Subsequently, the patient underwent a revision due to the dislocation of the humeral head and glenoid.While the surgeon was explanting the implants it was noted that the anterior ligament was torn and while removing the humeral head from the patient it disassociated from the taper implant.
 
Additional Manufacturer Narrative
(b)(4).D10: medical products: item#: 113042, versa-dial 46x18x53 hum head; lot#: j7283856.Item#: sagl2043, size 3 4-peg modular cemented glenoid; lot#: 65476156.H3: customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay corrected information.The following sections were updated: b4; b5; d2; g2; g3; g6; h1; h2 upon reassessment of the reported event, it was determined that the product was involved in the reported event.The initial report forwarded in error and should be voided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Additional Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional and corrected information.The original supplemental report was voided.After the completion of the investigation, it was determined that the original initial report should not have been voided and is reportable.The following sections were corrected: d9.Visual examination of the returned product/provided pictures identified the taper.The taper was returned attached to the humeral head as well.Review of the device history record(s) identified no deviations or anomalies during manufacturing related to the reported event.The rotator cuff is a group of muscles and tendons that surround the shoulder joint, keeping the head of the humerus firmly within the shallow socket of the shoulder.Tearing or stretching these supporting structures can contribute to subluxation/dislocation events.Typically, an anatomic shoulder is performed when the rotator cuff is stable and intact, and a reverse shoulder is performed for an insufficient rotator cuff.The root cause of the reported issue is not attributed to the reported device.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSA-DIAL/COMP TI STD TAPER
Common Device Name
PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER, UNCEMENTED/EXTREMITIES
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key20598903
Report Number0001825034-2024-02577
Device Sequence Number15689255
Product Code MBF
UDI-Device Identifier00880304217249
UDI-Public(01)00880304217249(17)320617(10)J7273860
Combination Product (Y/N)N
Initial Reporter StateMT
Initial Reporter CountryUS
PMA/510(K) Number
K193038
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/09/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number118001
Device Lot NumberJ7273860
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/07/2024
Supplement Date Received by Manufacturer11/05/2024
03/28/2025
Initial Report FDA Received Date11/04/2024
Supplement Report FDA Received Date12/03/2024
04/10/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/17/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11 NARRATIVE.
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age60 YR
Patient SexFemale
-
-