MENTOR TEXAS MENTOR MEMORYGEL XTRA BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
|
Back to Search Results |
|
| Catalog Number |
SMHX255 |
| Medical Device Problem Code |
Migration (4003)
|
| Health Effect - Clinical Code |
Deformity/ Disfigurement (2360)
|
| Date of Event |
09/10/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Device evaluation summary: according to the information received, the patient experienced migration of the breast implant.The investigation was completed on a photo of the suspect medical device because the physical device has not been received by mentor.Upon visual evaluation of the image provided in the complaint, no apparent damage or visual anomalies were observed.As the product involved in this complaint was not received, the device couldn't be analyzed according to our procedures.Implant displacement/migration may occur from improper implant sizing and/or placement, i.E., when the implant is too large or the pocket too small or when there has been inadequate preoperative assessment of stresses causing movement of the prosthesis.Implant displacement/migration is a known complication associated with these devices and is referenced in our current product insert data sheet.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Since no malfunction was observed during the investigation, no corrective and preventive action (capa) is required now.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.Reason for device explant and/or reoperation: implant migration.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent a breast augmentation procedure with a mentor memorygel xtra breast implant 255cc gel breast prosthesis that migrated post implantation.Migration of the patient¿s right breast prosthesis was reported.As a result, the patient underwent explantation on (b)(6) 2024.
|
| |
|
Additional Manufacturer Narrative
|
|
On 10-jan-2025, the mentor failure analysis lab received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.Additionally, a patient identifier (b)(6) was reported.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On 06-feb-2025, mentor completed the investigation on the suspect medical device.Device evaluation summary: according to the information received, the patient experienced implant displacement.The product was returned to mentor for evaluation.Mentor then conducted visual inspection of the returned device.During visual evaluation no apparent damage or visual anomalies were observed on the returned device.Implant displacement/migration may occur from improper implant sizing and/or placement, i.E., when the implant is too large or the pocket too small or when there has been inadequate preoperative assessment of stresses causing movement of the prosthesis.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Since no malfunction was observed during the investigation, no corrective and preventive action (capa) is required now.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.Manufacturer¿s reference number: (b)(4).
|
| |
|