• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE PKG.FIBEROPTIC CABLE 5MM X 10FT 3.05M; LIGHT SOURCE, FIBEROPTIC, ROUTINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE PKG.FIBEROPTIC CABLE 5MM X 10FT 3.05M; LIGHT SOURCE, FIBEROPTIC, ROUTINE Back to Search Results
Catalog Number 0233050084
Medical Device Problem Code Output Problem (3005)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/09/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that there was loss of light.
 
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
 
Additional Manufacturer Narrative
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: no light output.Probable root cause: broken optical fibers, poor light cable alignment in jaw, residue on fiber tip, nonuniform light output.The reported failure mode will be monitored for future reoccurrence.
 
Event or Problem Description
It was reported that there was loss of light.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PKG.FIBEROPTIC CABLE 5MM X 10FT 3.05M
Common Device Name
LIGHT SOURCE, FIBEROPTIC, ROUTINE
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
benjamin ly
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key20602037
Report Number0002936485-2024-00754
Device Sequence Number7458334
Product Code FCW
UDI-Device Identifier07613327051629
UDI-Public07613327051629
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K082813
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/03/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number0233050084
Device Lot Number165974
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/09/2024
Supplement Date Received by Manufacturer10/09/2024
Initial Report FDA Received Date11/04/2024
Supplement Report FDA Received Date01/03/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-