(b)(4).D10: medical product: vanguard ps tib brg 14x71/75mm: catalog#183644, lot#073790; van ps open intl fem-lt 60: catalog#183124, lot#002980; palacos mv+g 2x40: catalog#66031995, lot#91314797.G2: foreign: germany.The product will not be returned to zimmer biomet for investigation, as the product currently remains implanted.The investigation is in process.Upon completion of the investigation, a follow-up mdr be submitted.
|
|
It was reported that a patient underwent an initial left total knee arthroplasty.Subsequently, the patient presented with pain and instability.Imaging showed a quadriceps tendon rupture and articular surface fracture.Approximately five (5) years and six (6) months post-implantation, the patient underwent revision surgery where it was noted that a rotational error of the tibial component was causally responsible for the fractured bearing.The surgeon opted to retain the well-fixed tibial and femoral components due to the patient's advanced aged and a larger bearing was inserted.An initial patellar component was also implanted and the quadriceps tendon rupture was repaired with anchors.All available information has been reported.
|
|
This follow-up report is being submitted to relay additional information.The following sections have been updated: b4; b5; d4; g3; g4; h2; h3; h4; h6; h11.Visual examination of the returned product identified signs of implantation in the form of pits, gouges, micromotion and the post feature has fractured off the body of the implant.Dimensional analysis of the product determined that the product, where measured, was conforming to print specifications.The device was submitted for further analysis.Analysis determined the area on the main body of the bearing with the remnant of the base of the post shows evidence of possible impingement damage from the femoral component; however, there is no similar impingement damage on the post fragment due to the apparent loss of material in that area.No additional fragments were returned for examination, however.The post fragment has several additional areas showing possible evidence of abrasions or wear.The fractured post was caused by overloading, possibly exacerbated by impingement by the femoral component.Device history record was reviewed and no discrepancies related to the reported event were found.Medical records were provided and reviewed by a health care professional.Review of the available records identified the following: from revision report: significant instability in the case of inlay fracture, quadriceps tendon rupture, pronounced retropatellar arthrosis, patient presented with clearly unstable and painful left knee rom -10-90° with canes, typical retropatellar symptoms.Per surgeon: a rotational error of the tibia is causally responsible for the broken inlay.In the case of advanced age, decision to keep the fixed implants.Investigation results concluded that the reported event was due to failure of the operator to follow instructions.Review of the provided medical records found that it is noted that the tibial implant is rotated and in an incorrect position.Per the ifu, improper selection, placement, positioning, alignment and fixation of the implant components may result in unusual stress conditions which may lead to subsequent reduction in the service life of the prosthetic components.Malalignment of the components or inaccurate implantation can lead to excessive wear and/or failure of the implant or procedure.A summary of the investigation was sent to the complainant conveying proper surgical technique and use of the device.Reported event was confirmed by review of provided medical records and product return.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|