| Model Number |
URF-V3 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device evaluation is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was observed during the device inspection, that the uretero reno videoscope exhibited deformation inside the instrument channel.There were no reports of patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.New information was added to the following fields: d8, h3, h4, and h6.Corrected data: d9.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, olympus believes that the deformation inside the instrument channel could have occurred if the device had been subjected to physical stress.Olympus will continue to monitor field performance for this device.
|
| |
|