MAUDE Adverse Event Report: RESPIRONICS, INC. BIPAP A40 SYS SILVER SERIES FLOW GEN; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. BIPAP A40 SYS SILVER SERIES FLOW GEN; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE Back to Search Results
Model Number 1109601
Medical Device Problem Code Electrical /Electronic Property Problem (1198)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/20/2018
Type of Reportable Event Malfunction
Event or Problem Description
The manufacturer received information alleging a ventilator inoperative condition occurred.There was no harm or injury reported.The device was sent to a service center for evaluation.The ventilator inoperative condition was not duplicated.The device's sd card was found to require replacement, which could cause device errors.The unit was not repaired as it was scrapped.Prior to identifying the issue described in fsn 2023-cc-src-039, complaints associated with the issue did not meet requirements for vigilance reporting as a reportable incident.As part of the hhe process, a retrospective review of complaints associated with the issue described in fsn 2023-cc-src-039 was necessary to reassess reportability.Upon this further assessment, these complaints are being reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIPAP A40 SYS SILVER SERIES FLOW GEN
Common Device Name
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
melissa rosko
6501 living place
pittsburgh, PA 15208
4125423300
MDR Report Key20606098
Report Number2518422-2024-67767
Device Sequence Number18125061
Product Code MNT
UDI-Device Identifier00606959042797
UDI-Public(01)00606959042797
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K121623
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 11/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number1109601
Device Catalogue Number1109601
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/17/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/05/2018
Initial Report FDA Received Date11/05/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/05/2015
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1813-2024
Patient Sequence Number1
Patient SexUnknown
-
-