| Model Number |
9733560 |
| Medical Device Problem Codes |
Display or Visual Feedback Problem (1184); Application Program Freezes or Fails to Launch (4031)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/31/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H3, h6: no products have been returned to medtronic for analysis.Codes b17, c20, and d15 are applicable.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Medtronic received information regarding a navigation system being used outside of a procedure.It was reported that the screen no longer received a signal.There was no video and a black screen.The probable cause was suspected to likely be the computer or complementary metal oxide semiconductor (cmos) battery.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
Continuation of d10: section d information references the main component of the system.Other relevant device(s) are: product id: 9735672, serial lot/sw version: - img code g02007 applies to the complementary metal-oxide semiconductor (cmos) battery (cmos) battery.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|