| Model Number |
WA2UR31A |
| Medical Device Problem Code |
Loose or Intermittent Connection (1371)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/17/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device evaluation is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was observed during the device inspection, that the ureterscope exhibited a loose stop cock.There were no reports of patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.New information was added to the following fields: d8, h3, and h6.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, olympus presumes that the damaged rotary valve can be attributed to improper handling and application of excessive force, impact to the device like a fall, shock, or similar stress.Olympus will continue to monitor field performance for this device.
|
| |
|