MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH URETEROSCOPE 6.7/8.4 FR. X 430 MM, 0°, ANGLED OCULAR, 4.2 FR. CHANNEL; WA2UR31A

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH URETEROSCOPE 6.7/8.4 FR. X 430 MM, 0°, ANGLED OCULAR, 4.2 FR. CHANNEL; WA2UR31A Back to Search Results
Model Number WA2UR31A
Medical Device Problem Code Loose or Intermittent Connection (1371)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/17/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device evaluation is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
It was observed during the device inspection, that the ureterscope exhibited a loose stop cock.There were no reports of patient involvement.
 
Additional Manufacturer Narrative
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.New information was added to the following fields: d8, h3, and h6.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, olympus presumes that the damaged rotary valve can be attributed to improper handling and application of excessive force, impact to the device like a fall, shock, or similar stress.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
URETEROSCOPE 6.7/8.4 FR. X 430 MM, 0°, ANGLED OCULAR, 4.2 FR. CHANNEL
Common Device Name
WA2UR31A
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key20606967
Report Number9610773-2024-02646
Device Sequence Number20874021
Product Code FGB
UDI-Device Identifier04042761086561
UDI-Public04042761086561
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K200369
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/03/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberWA2UR31A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/20/2024
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/25/2024
Supplement Date Received by Manufacturer02/14/2025
Initial Report FDA Received Date11/05/2024
Supplement Report FDA Received Date03/03/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/08/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-