The reported event was inconclusive because no sample was returned.The product was used for urological care.It was unknown whether the product had caused the reported failure.However, the potential root cause for this failure mode could be the catheters was contact with sharp object or exposure to petrolatum based products.A review of the device history record did not show any problems or condition that would have contributed to the reported issue.The instructions for use were found adequate and state the following: structure and composition: the product include three series: bardex, bardic and bardia.The composition of each series is as follows: bardex, bardic and bardia bi-lumen series are composed of catheter body, inflation cone interface, balloon and valve.Bardia tri-lumen is composed to tube body, inflation cone interface, deflation cone interface, flush cone interface, balloon and valve.The product is with silicone coating.The material of catheter body, inflation conical interface, deflation conical interface, flushing conical interface, balloon is natural latex.The material of valve is polypropylene.Sterilized by gamma radiation.Disposable.Scope of application: foley catheter is intended for use in the drainage of urine from the bladder of children and adults.Precaution: this product contains natural rubber latex which may cause allergic reaction.Sterile unless package has been opened or damage.Warning: do not use petroleum substrate lubricants with the latex- based urethral catheters such as petroleum jelly and label liquid paraffin, which will damage latex and may burst balloon.Water-soluble lubricant can be used.Valve type: use luer slip syringe.Do not use needle.To deflate catheter balloon: gentle insert a syringe in the catheter valve.Never use excessive force to make the syringe stick in the valve.If you notice slow or no deflation, re-seat the syringe gently.Use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen.If permitted by hospital practical, the valve may be cut off.If this fails, please contact an adequately trained professional for assistance, as directed by hospital practical.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient.Storage: catheters need to be stored at room temperature and keep away from the direct light and preferably stored in the original packing box.Do not use the product of have latex allergy.This report references a us equivalent device.The us udi equivalent for this product number is used.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
It was reported that the patient was admitted to the hospital due to "right hip injury and pain caused by a fall, accompanied by difficulty in movement for 1 day".The diagnosis upon admission: right intertrochanteric fracture.At noon on (b)(6) 2024, the patient was unable to urinate on their own, so they were given an indwelling catheter.15 ml of saline was injected as a fixative.The operation went smoothly, and the drainage of urine was unobstructed.The patient's family discovered that the foley catheter had fallen off on its own at 18:40 on (b)(6), so they notified the nurse on duty, disinfected the external urethral opening, and conducted clinical medical observation.This incident did not cause any impact on the patient.The doctor and nurse on duty examined the fallen catheter and found that the catheter water storage bag was ruptured.Since the patient could already urinate on his own, there was no need to re-place the catheter.
|