| Catalog Number |
100/870/080CZ |
| Medical Device Problem Codes |
Break (1069); Crack (1135)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/01/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the cannula was cracked and breaking.This occurred with 2 cannulas.There was patient involvement, and no patient harm/adverse event reported.
|
| |
|
Additional Manufacturer Narrative
|
|
B3.Date of event: month and year of event have been provided, but day is unknown.H3.Reason device not evaluated by mfg: other; device was not returned to manufacturer.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: no product sample was received.Investigation of this complaint was done based on photos.On the photos there is shown one 005/073/680 9mm blu thin wall inner cannula with visible crack as described by customer.Affected inner cannula is disposable device which is regularly cleaned by customer and therefore it shall be replaced based on its condition as described in ifu (instructions for use).To verify blue line ultra-thin wall inner cannulas durability, we carried out informative testing on randomly selected inner cannula samples of each size (6.0mm - 10.0mm, both fenestrated and non-fenestrated versions) from current batches to measure their performance using established standard test methods.Based on results of the testing, current blu thin wall inner cannulas were found to be suitable for its function.Unfortunately, without the sample / more information we are unable to further assess this incident nor further investigate.Investigation might continue if sample or more information is provided.A review of the device history record (dhr) shows there were no observations or nonconformities recorded during manufacture to suggest an issue of this nature would occur with this lot of products.
|
| |
|
Search Alerts/Recalls
|