MAUDE Adverse Event Report: COVIDIEN MEDICAL PRODUCTS PREMIUM PLUS CEEA; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN MEDICAL PRODUCTS PREMIUM PLUS CEEA; STAPLE, IMPLANTABLE Back to Search Results
Model Number 111985
Medical Device Problem Codes Break (1069); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/15/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
According to the reporter, during a laparoscopic rectal resection, during rectal anastomosis, when firing, the surgeon felt as if the device was dislodged during needle insertion.Giving the impression that it has not been fired firmly.The wing nut was turned twice to tilt the anvil, but since it was turned more than two times, the anvil came off when the anastomosis device was removed.The donut ring was confirmed to be complete so the operation was completed with the same stapler.There was no patient injury.
 
Additional Manufacturer Narrative
Correction: g3, h6.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
H3: evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that the anvil disengaged and the wing nut was broken.The reported issues could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PREMIUM PLUS CEEA
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
COVIDIEN MEDICAL PRODUCTS
building 10- no 789 puxing roa
shanghai 20111 4
CH  201114
Manufacturer (Section G)
COVIDIEN MEDICAL PRODUCTS
building 10- no 789 puxing roa
shanghai 20111 4
CH   201114
Manufacturer Contact
justin ellis
8200 coral sea st ne
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key20610593
Report Number3006451981-2024-00153
Device Sequence Number7928669
Product Code GDW
UDI-Device Identifier10884523006599
UDI-Public10884523006599
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K001895
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/08/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/11/2025
Device Model Number111985
Device Catalogue Number111985
Device Lot NumberS0CC001LX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/30/2024
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/18/2024
Supplement Date Received by Manufacturer11/12/2024
12/20/2024
Initial Report FDA Received Date11/05/2024
Supplement Report FDA Received Date11/13/2024
01/08/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/07/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-