• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. THREE PEG PATELLA 41MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. THREE PEG PATELLA 41MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Catalog Number 200-02-41
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 10/21/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
D10: concomitants: (b)(6) - gps implant kit v2 10041320190 02-010-04-0340 - logic cr femoral por, right, sz 4 (b)(6) 521-78-32 - threaded pin size 3.0 collarless (b)(6) 201-78-81 - 3 trocar, mod.Hex 2pk (b)(6) 201-78-15 - holding pin mini sharp point 4 pk (b)(6) 02-012-45-4050 - lgc tibial fit tray cem sz 4f / 5t (b)(6) 521-78-23 - threaded pin size 2.3 collared (b)(6) 521-78-23 - threaded pin size 2.3 collared (b)(6).
 
Event or Problem Description
It was reported that a 63 yo male patient who had an initial right knee implanted in (b)(6)2021, and was revised in (b)(6) 2022, underwent a second revision procedure in (b)(6) 2024, approximately 2 years post the (b)(6) 2022 procedure.Everything was revised due to wear on patella and tibial insert and was replaced with a competitor¿s devices.There were no device breakages or surgical delays during the procedure.The patient was last known to be in stable condition following the event.X-rays were provided.The explanted devices are not available for analysis.They were discarded at the hospital.Device images were provided.No further information.First revision reported under mdr# 1038671-2022-01234.
 
Additional Manufacturer Narrative
Report number: 1038671-2024-04238 this follow-up report is being submitted to relay additional and/or corrected information.The following sections were updated: h3: device not returned the following sections were corrected: b2 d1 h6 the reason for the revision may be the result of wear of the patella and third body wear on the tibial insert.Contributions of patient-related issues or implant alignment to the wear on the patella cannot be confirmed from the information provided and the devices were not available for evaluation.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THREE PEG PATELLA 41MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer Contact
geoff gannon
2320 nw 66 court
gainesville, FL 32653
3525541928
MDR Report Key20612763
Report Number1038671-2024-04239
Device Sequence Number18183652
Product Code JWH
UDI-Device Identifier10885862039620
UDI-Public10885862039620
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K932690
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/22/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/06/2024
Device Catalogue Number200-02-41
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/22/2024
Supplement Date Received by Manufacturer01/21/2025
Initial Report FDA Received Date11/05/2024
Supplement Report FDA Received Date01/22/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/09/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age63 YR
Patient SexMale
-
-