ALPHATEC SPINE, INC. INVICTUS; SCREWDRIVER Back to Search Results
Model Number 17950-225
Medical Device Problem Codes Break (1069); Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/25/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the tip of the driver broke off when inserting the screw into the ilium.All broken pieces were removed from the patient.No further complications were reported.This is report 2 of 2.
 
Additional Manufacturer Narrative
The device is currently being evaluated.A follow-up report with the results of the investigation will be submitted upon completion.
 
Additional Manufacturer Narrative
This is report 2 of 2.Corrected information: h3.Yes, h6.Investigation findings: 3252, investigation conclusion: 61.Device evaluation: visual inspection confirms that the distal tip has sheared off at the base of the hexalobe.This failure is likely due to the applied torsional force being greater than the yield strength.Review of device history records showed no manufacturing or processing related irregularities.Labeling review: warnings/cautions/precautions: risks identified with the use of these devices, which may require additional surgery, include device component failure, loss of fixation/stabilization, non-union, vertebral fracture, neurological injury, vascular or visceral injury.Possible adverse effects: initial or delayed loosening, disassembly, bending, dislocation, and/or breakage of device components.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INVICTUS
Common Device Name
SCREWDRIVER
Manufacturer (Section D)
ALPHATEC SPINE, INC.
1950 camino vida roble
carlsbad CA 92008
Manufacturer (Section G)
ALPHATEC SPINE, INC.
1950 camino vida roble
carlsbad CA 92008
Manufacturer Contact
wesley channell
1950 camino vida roble
carlsbad, CA 92008
9014283693
MDR Report Key20613048
Report Number2027467-2024-00141
Device Sequence Number8874412
Product Code HXX
UDI-Device Identifier00190376473017
UDI-Public(01)00190376473017(10)EM53910
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 11/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number17950-225
Device Catalogue Number17950-225
Device Lot NumberEM53910
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/30/2024
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/08/2024
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date11/05/2024
Supplement Report FDA Received Date11/12/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/29/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-