Model Number CYF-VHA
Medical Device Problem Code Failure to Clean Adequately (4048)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device evaluation is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
It was observed during device evaluation that the cysto-nephro videoscope had foreign material on the adhesive around the objective lens.There was no patient involvement.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, olympus confirmed foreign material came out of the device, but the type of the material or root cause cannot be identified.A definitive root cause cannot be determined.The event can be detected/prevented by following the instructions for use which state: there is a reprocessing manual for cyf-vha, and it describes the reprocessing procedures.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CYSTO-NEPHRO VIDEOSCOPE
Common Device Name
CYSTO-NEPHRO VIDEOSCOPE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key20613133
Report Number3002808148-2024-10796
Device Sequence Number12645945
Product Code FAJ
UDI-Device Identifier04953170411274
UDI-Public04953170411274
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K062049
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 11/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberCYF-VHA
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/21/2024
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/23/2024
Supplement Date Received by Manufacturer11/14/2024
Initial Report FDA Received Date11/05/2024
Supplement Report FDA Received Date11/15/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/11/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-