| Catalog Number |
71674067 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/10/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: (b)(4).Two lot numbers were provided, 21acw3013 and 18acw1007, but it has not been confirmed which one corresponds to the fractured screwdriver.If any additional information is received, it will be sent in a supplemental report.
|
| |
|
Event or Problem Description
|
|
It was reported that, during an orif revision surgery, one (1) lag screw driver fractured when trying to thread the driver onto the lag screw.The broken piece was accounted for immediately.It was still attached to the driver when it happened, so it came off and didn¿t need to be retrieved.This occurred at a complex broken compression screw removal.The procedure was resumed, without any delay, using a s+n backup device.No injury was reported as a result of this issue.
|
| |
|
Additional Manufacturer Narrative
|
|
D4, h4: updated.
|
| |
|
Additional Manufacturer Narrative
|
|
H3, h6: the associated device was returned and evaluated.The visual inspection revealed that a piece of the threaded tip broke off.This piece was not returned.The lag screw driver portion of the device was not returned as well.This inspection was conducted by naked eye.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history revealed similar events for the listed part number over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no evidence to conclude that the product failed to meet any specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely potential factors that could contribute to the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|
Search Alerts/Recalls
|
|
|