| Model Number |
0636 |
| Medical Device Problem Code |
Over-Sensing (1438)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/09/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that this right ventricular (rv) lead was attempted due to oversensing.A new lead with same model was implanted.No adverse patient effects were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that this right ventricular (rv) lead was attempted due to oversensing.A new lead with same model was implanted.No adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon receipt at our post market quality assurance laboratory, a thorough evaluation of the lead was performed.Testing was completed to assess lead electrical performance and inner/outer insulation integrity.Measurements throughout these tests were within normal limits.Microscopic inspections of the terminal pin assembly, lead body, and electrode tip found no anomalies.Laboratory testing was unable to reproduce the reported clinical observations and detailed analysis did not reveal any abnormalities.Typical surgery-related damage was noted but is not considered abnormal.
|
| |
|