It was reported the physician was implanting a 37mm gore® cardioform asd occluder to close an atrial septal defect.The device could not fully capture the superior rim, so it was decided to remove the device.The angle made it difficult to recapture the device, and it caught on the tip of the delivery system, breaking the retrieval cord.The device embolized into the right ventricle.The device was then snared out of the pulmonary artery.A new 44mm gore® cardioform asd occluder was implanted with no issues.The patient was doing well following the procedure.
|
|
Per the gore® cardioform asd occluder instruction for use, in the section ¿potential device ¿ or procedure-related adverse events¿.Adverse events associated with the use of the occluder may include, but are not limited to: device embolism.A4: no patient weight available.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|