• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MENTOR TEXAS MENTOR CPX 4 BREAST TISSUE EXPANDER WITH SUTURE TABS; EXPANDER, SKIN, INFLATABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MENTOR TEXAS MENTOR CPX 4 BREAST TISSUE EXPANDER WITH SUTURE TABS; EXPANDER, SKIN, INFLATABLE Back to Search Results
Catalog Number 3549213
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Deformity/ Disfigurement (2360)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
On november 28, 2024, the mentor failure analysis lab received the device for evaluation.The analysis has begun, but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: left deflation mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
 
Event or Problem Description
It was reported that a 40-year-old caucasian female patient underwent breast surgery with 450cc mentor cpx 4 breast tissue expander with suture tabs and experienced left side expander deflation.As a result, the patient underwent replacement surgery with unknown mentor devices on (b)(6) 2024.
 
Additional Manufacturer Narrative
On january 14, 2025, device evaluation was completed as follow: mentor conducted a visual inspection, leak testing, and microscopic examination of the returned device.During the visual analysis of the returned sample, no apparent damage or visual anomalies were observed.Leak testing was performed, according to the mentor procedure, and it revealed a tear between the union of the shell and the suture tab at the 3 o'clock position.No other leak sites were found during the analysis.A microscopic examination was performed, and the cause of the tear could not be identified.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Based on the information currently available, a notification was sent to mentor's manufacturing team for further investigation.According to the manufacturing investigation, potential process failures resulting in a tear that may represent deflation were assessed regarding the product complaint.According to the manufacturing investigation, the collected process controls and data records for the deflated tissue expander meet specifications.The deflation mentioned in the product complaint cannot be conclusively confirmed as a manufacturing-related defect.Although no conclusion could be reached on the cause of the reported event, the instructions for use contain the following caution: the silicone shell of the tissue expander may easily be cut by a scalpel or ruptured by excessive stress, manipulation with blunt instruments, or penetration by a needle.Subsequent deflation and/or rupture could result.All devices should be carefully inspected for structural integrity prior to and during implantation.In addition, the patient should be advised that vigorous body movement (e.G., physical exercise) or excessive manipulation or trauma in the expander region may cause stress to the device and result in subsequent deflation.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.No corrective and preventive action (capa) is required now.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post-market surveillance.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR CPX 4 BREAST TISSUE EXPANDER WITH SUTURE TABS
Common Device Name
EXPANDER, SKIN, INFLATABLE
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
kate karberg
3041 skyway circle north
irving, TX 75038
3035526892
MDR Report Key20618139
Report Number1645337-2024-12709
Device Sequence Number12625801
Product Code LCJ
UDI-Device Identifier00081317006732
UDI-Public(01)00081317006732(17)280402(10)9988509
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K130813
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number3549213
Device Lot Number9988509
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/24/2024
Supplement Date Received by Manufacturer11/28/2024
01/14/2025
Initial Report FDA Received Date11/06/2024
Supplement Report FDA Received Date12/13/2024
01/14/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/04/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age40 YR
Patient SexFemale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-