| Model Number |
MI1210 SYNCHRONY ST |
| Medical Device Problem Code |
Impedance Problem (2950)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
10/31/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Event or Problem Description
|
|
The user's hearing performance with the device is affected after a trauma to the head.Re-implantation is considered.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: damage to the active electrode which is consistent with minute device mobility was determined to have led to device failure over time due to fatigue breakages.Also, the reported impact to the head might have weakened the fixation of the active electrode and could also be a factor for electrode mobility.The investigation results appear to match the problems mentioned in the recipient report.This is a final report.
|
| |
|
Event or Problem Description
|
|
Reportedly the user fell from a chair and hit her left implant site on (b)(6) 2024.Immediately after the hit, she stopped hearing with the device.The user has been re-implanted.
|
| |
|