MEDTRONIC SOFAMOR DANEK USA, INC ZEVO¿ ANTERIOR CERVICAL PLATE SYSTEM; SCREWDRIVER Back to Search Results
Model Number 3030007
Medical Device Problem Codes Break (1069); Naturally Worn (2988)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/08/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product analysis: part # 3030007, lot # im20c011 visual inspection confirmed the tip of the driver is worn from repeated use.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
Information was received from a manufacturer representative (quality inspector) regarding a part used for spinal therapy.It was reported that the tip was broken and instrument was worn.There was no further information reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZEVO¿ ANTERIOR CERVICAL PLATE SYSTEM
Common Device Name
SCREWDRIVER
Manufacturer (Section D)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
Manufacturer (Section G)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
Manufacturer Contact
justin ellis
1800 pyramid place
memphis, TN 38132
7635265677
MDR Report Key20619885
Report Number1030489-2024-01444
Device Sequence Number12645855
Product Code HXX
UDI-Device Identifier00643169342293
UDI-Public00643169342293
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 11/06/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number3030007
Device Catalogue Number3030007
Device Lot NumberIM20C011
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/31/2024
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/09/2024
Initial Report FDA Received Date11/06/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/16/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1