| Model Number |
8887605189 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Customer reports: the balloon on the catheter snapped (burst).Sometimes immediately after insertion, other times 2-3 days after insertion.The end-user's urine is clean and has no grit.Three occurrences were reported for lot number 23b097fhx, this case represents #3 of the 3 reported cases.The balloon was removed as per normal procedure and the catheter has been replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
An investigation is currently underway.Upon completion, the results will be forwarded.H3: device narrative corrected data regulatory attachments narrative corrected data regulatory attachments evaluation comment - the customer stated that the device will not be returned for evaluation but did not provide rationale.
|
| |
|
Additional Manufacturer Narrative
|
|
The device history record for 23b097fhx was reviewed and no anomalies related to the reported issue were observed.There were no photos or physical samples received for evaluation.Because a sample was not returned, we were unable to perform a follow up investigation to include functional and visual evaluations to confirm the issue and root cause analysis.At this time, a corrective and preventive action is not deemed necessary.This information will be used for tracking and trending purposes, and we will continue to monitor.Section d4 (unique identifier (udi) #): originally reported as (b)(4).Section d2b product code: originally reported as ezl and should be fcm.Section e1: please disregard the originally reported phone number.The phone number provided was invalid and the customer phone number is unknown.
|
| |
|
Search Alerts/Recalls
|