|
This follow-up report is being submitted to relay additional information.The following sections were updated: b4, b5, d4, g3, g6, h2, h3, h4, h6, h11.Visual examination of the provided pictures identified signs of use (scratches and nicks).The device fractured at the weld location.No device returned; no further analysis can be made.Review of the device history records identified no deviations or anomalies during manufacturing.A definitive root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.This complaint can be confirmed via the provided picture.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|