| Model Number |
OMS-T10BTNL |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/22/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
According to the reporter, during the laparoscopic hernia repair, the surgeon was inserting mesh through the trocar when the valve seal inside the patient detached and stuck to the mesh.The surgeon was able to retrieved it with no issues.It was also noted that when the surgeon was cleaning the second trocar with a port cleaner, the tip of the cleaner got stuck in the trocar.A new trocar was opened and used.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
D10 concomitant medical product: oms-t10btnl, omst10btnl 10 blunt tip trocar w/o latex (lot#p4f1046) medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted that the main valve seal was cut and disengaged.The 5sd converter door was disengaged.The trocar balloon, lock collar and 7/8 converter doors was intact.Functionally, the lock collar moved up and down the cannula and securely locked in place.The 7/8 converter door moved properly and was fixed securely in place.The balloon was inflated using a test syringe, no leaks detected.It was reported that the seal was disengaged.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.This issue may occur when excessive force is applied during clinical application.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: damage to the balloon by instruments used during insertion and in the course of a procedure may cause device failure.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|