Model Number 02.07.0214SCF
Medical Device Problem Code Device Appears to Trigger Rejection (1524)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 10/04/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Batch review performed on 11-oct-2024: lot 2011833: (b)(4) items manufactured and released on 04-mar-2021.Expiration date: 2026-02-15.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold with no similar reported event during the period of review.Infection is a known possible complication of any surgery.Although no root cause can be established, there is no indication that any issue with the device may have caused or contributed to the event, and the document review does not indicate any potential manufacturing related cause.
 
Event or Problem Description
The patient undergone a primary surgery with competitor implants.On the (b)(6) 2024, the patient has been revised and gmk-revision components have been implanted.On the (b)(6) 2024, the patient came in due to signs of an infection and the pathogen is unknown.The surgeon performed a washout and revised the liner.The surgery was completed successfully.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GMK REVISION TOTAL KNEE SYSTEM
Common Device Name
KNEE TIBIAL INSERT
Manufacturer (Section D)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, switzerland 6874
SZ  6874
Manufacturer (Section G)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, switzerland 6874
SZ   6874
Manufacturer Contact
marco giannessi
strada regina
castel san pietro, switzerland 6874
SZ   6874
MDR Report Key20621603
Report Number3005180920-2024-00899
Device Sequence Number7353898
Product Code JWH
UDI-Device Identifier07630030816710
UDI-Public07630030816710
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K103170
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 11/06/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/15/2026
Device Model Number02.07.0214SCF
Device Lot Number2011833
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/07/2024
Initial Report FDA Received Date11/06/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/04/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-