| Catalog Number |
731M0000 |
| Medical Device Problem Code |
Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/14/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Current information is insufficient to permit a valid conclusion about the cause of this event.A follow up report will be sent upon completion of the device evaluation.Reference report 3012447612-2024-00307.
|
| |
|
Event or Problem Description
|
|
It was reported that two vitality driver tips from a power set became twisted intra-operatively.The procedure was completed without any patient impact.This is report two of two for this event.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: "device evaluation anticipated, but not yet begun" is erroneous and no longer applies.The device was evaluated.Corrections in h3.Additional information in h6: investigation type, findings, and conclusions.Device evaluation: the returned device matches the information in the complaint file and was examined.Visual inspection revealed signs of wear on the tip (ie: rounding of the hex edges).A functional inspection was performed on the driver with the worn tip with a vitality screw (07.02000.075 lot t09180) and found it was able to engage and disengage as expected.Root cause: a definitive root cause cannot be determined with the information provided.This event could possibly be attributed to wear through use over time.Dhr review: the dhr was reviewed.There are no indications of manufacturing issues which would have contributed to this event and the device was likely conforming when it left highridge medical¿s control.Device usage: this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
|
| |
|
Event or Problem Description
|
|
It was reported that two vitality driver tips from a power set became twisted intra-operatively.The procedure was completed without any patient impact.This is report two of two for this event.
|
| |
|