MAUDE Adverse Event Report: EXACTECH, INC. OPTETRAK TOTAL KNEE SYSTEM, HI-FLEX ASYMMETRIC POSTERIOR-STABILIZED CEMENTED FEM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. OPTETRAK TOTAL KNEE SYSTEM, HI-FLEX ASYMMETRIC POSTERIOR-STABILIZED CEMENTED FEM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Catalog Number 02-010-01-0250
Medical Device Problem Codes Unstable (1667); Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Codes Pain (1994); Synovitis (2094); Discomfort (2330); Osteolysis (2377); Joint Laxity (4526); Implant Pain (4561); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 12/01/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device was not returned for evaluation and no medical or other records containing treatment information or patient information have been received; therefore, the reported event cannot be confirmed, nor can the circumstances or potential causes or contributors to the alleged event be evaluated.Should additional, material information become available that permits more analysis or conclusions, a supplemental report will be filed accordingly.
 
Event or Problem Description
It was reported that approximately 128 months after a left total knee replacement procedure, the patient underwent a revision procedure to address loosening, painful left knee implant and instability.Revision surgery operative notes were provided.Patient left the operating room in stable condition.Post operative diagnosis noted wear of the implant.No further issues or complications were reported.No additional information is available.
 
Additional Manufacturer Narrative
D1: correct: d10 concomitants: (b)(6), 208-25-13 - cc tibial insert sz 5, 13mm, (b)(6), 200-02-38 - three peg patella 38mm, (b)(6), 02-012-41-5040 - logic tibia traptray cem sz 5f/4t, (b)(6), 02-012-41-5050 - logic tibia traptray cem sz 5f/5t, (b)(6), 02-012-41-5050 - logic tibia traptray cem sz 5f/5t, (b)(6), 204-04-55 - trapezoid tibial tray sz 5f/5t, (b)(6), 204-32-01 - fluted stem extension 11l x 12 mm, (b)(6), 02-012-35-5009 - logic tibia ps mod insrt sz 5 9mm, (b)(6), 200-02-38 - three peg patella 38mm, (b)(6), 02-010-01-0250 - logic femoral ps cem left sz 5, (b)(6), 02-012-44-5011 - logic tibia implant psc insert, sz 5, 11mm, (b)(6), 02-012-44-5011 - logic tibia implant psc insert, sz, (b)(6), 200-02-35 - three peg patella 35mm, (b)(6), 200-02-38 - three peg patella 38mm, (b)(6), 210-01-05 - nmc femoral sz 5.H6: corrected health effect - clinical code, medical device problem code, type of investigation.The reason for the revision reported cannot be confirmed from the information provided but may be the result of prosthesis wear, instability, and femoral loosening or due to inclusion of the polyethylene in the packaging recall.Additionally, these devices appear to have been implanted for over ten years prior to the reported event.Potential contributions of manufacturing, user, and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images, radiographs, and product information were not provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTETRAK TOTAL KNEE SYSTEM, HI-FLEX ASYMMETRIC POSTERIOR-STABILIZED CEMENTED FEM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC
2320 nw 66th ct.
gainesville FL 32653
Manufacturer Contact
geoff gannon
2320 nw 66th ct
gainesville, FL 32653
3523771140
MDR Report Key20623099
Report Number1038671-2024-04271
Device Sequence Number15291991
Product Code JWH
UDI-Device Identifier10885862001177
UDI-Public10885862001177
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K033883
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/02/2016
Device Catalogue Number02-010-01-0250
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/01/2023
Supplement Date Received by Manufacturer11/15/2024
Initial Report FDA Received Date11/06/2024
Supplement Report FDA Received Date11/19/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/05/2011
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11.
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age67 YR
Patient SexMale
-
-