D1: correct: d10 concomitants: (b)(6), 208-25-13 - cc tibial insert sz 5, 13mm, (b)(6), 200-02-38 - three peg patella 38mm, (b)(6), 02-012-41-5040 - logic tibia traptray cem sz 5f/4t, (b)(6), 02-012-41-5050 - logic tibia traptray cem sz 5f/5t, (b)(6), 02-012-41-5050 - logic tibia traptray cem sz 5f/5t, (b)(6), 204-04-55 - trapezoid tibial tray sz 5f/5t, (b)(6), 204-32-01 - fluted stem extension 11l x 12 mm, (b)(6), 02-012-35-5009 - logic tibia ps mod insrt sz 5 9mm, (b)(6), 200-02-38 - three peg patella 38mm, (b)(6), 02-010-01-0250 - logic femoral ps cem left sz 5, (b)(6), 02-012-44-5011 - logic tibia implant psc insert, sz 5, 11mm, (b)(6), 02-012-44-5011 - logic tibia implant psc insert, sz, (b)(6), 200-02-35 - three peg patella 35mm, (b)(6), 200-02-38 - three peg patella 38mm, (b)(6), 210-01-05 - nmc femoral sz 5.H6: corrected health effect - clinical code, medical device problem code, type of investigation.The reason for the revision reported cannot be confirmed from the information provided but may be the result of prosthesis wear, instability, and femoral loosening or due to inclusion of the polyethylene in the packaging recall.Additionally, these devices appear to have been implanted for over ten years prior to the reported event.Potential contributions of manufacturing, user, and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images, radiographs, and product information were not provided.
|